Attributes
attributes.py - Contains the ID of all attributes belongs to the module.
- class nirfmxgsm.attributes.AttributeID(value)[source]
Bases:
EnumThis enum class contains the ID of all attributes belongs to the module.
- AUTO_LEVEL_INITIAL_REFERENCE_LEVEL = 4247552
Specifies the initial reference level the
auto_level()method uses to estimate the peak power of the input signal. This value is expressed in dBm.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 30.
- AUTO_TSC_DETECTION_ENABLED = 4194323
Specifies whether the measurement automatically detects the training sequence code (TSC).
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is True.
Name (Value)
Description
False (0)
The measurement uses the value that you configure using the TSC attribute.
True (1)
The measurement detects the TSC in the burst.
- BAND = 4194318
Specifies the operation band.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is PGSM.
Name (Value)
Description
PGSM (0)
The operation band is Primary GSM in the 900 MHz band.
EGSM (1)
The operation band is Extended GSM in the 900 MHz band.
RGSM (2)
The operation band is Railway GSM in the 900 MHz band.
DCS1800 (3)
The operation band is GSM in the 1800 MHz band.
PCS1900 (4)
The operation band is GSM in the 1900 MHz band.
GSM450 (5)
The operation band is GSM in the 450 MHz band.
GSM480 (6)
The operation band is GSM in the 480 MHz band.
GSM850 (7)
The operation band is GSM in the 850 MHz band.
GSM750 (8)
The operation band is GSM in the 750 MHz band.
TGSM810 (9)
The operation band is terrestrial GSM in the 810 MHz band.
- BURST_SYNCHRONIZATION_TYPE = 4194326
Specifies the method used to synchronize the burst.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is TSC.
Name (Value)
Description
TSC (0)
Synchronizes the measurement to the timing of the demodulated training sequence in the burst. The measurement requires a burst with a valid training sequence code (TSC). The measurement determines the T0 point by demodulating the burst and identifying the TSC.
Amplitude (1)
Synchronizes the measurement based on the RF envelope of the received signal. The measurement sets the T0 point as the center of the RF envelope.
None (2)
Sets the T0 point to 273.23 microseconds after the trigger.
- BURST_TYPE = 4194321
Specifies the burst type.
Use “slot<n>” as the selector string to configure or read this attribute.
The default value is NB.
Name (Value)
Description
NB (0)
The burst type is Normal Burst.
HB (1)
The burst type is Higher Symbol Rate Burst.
AB (2)
The burst type is Access Burst.
- CENTER_FREQUENCY = 4194305
Specifies the expected carrier frequency of the acquired RF signal. This value is expressed in Hz. The signal analyzer tunes to this frequency.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default of this attribute is hardware dependent.
- DIGITAL_EDGE_TRIGGER_EDGE = 4194310
Specifies the active edge for the trigger. This attribute is used only when you set the
TRIGGER_TYPEattribute to Digital Edge.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Rising Edge.
Name (Value)
Description
Rising Edge (0)
The trigger asserts on the rising edge of the signal.
Falling Edge (1)
The trigger asserts on the falling edge of the signal.
- DIGITAL_EDGE_TRIGGER_SOURCE = 4194309
Specifies the source terminal for the digital edge trigger. This attribute is used only when you set the
TRIGGER_TYPEattribute to Digital Edge.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default of this attribute is hardware dependent.
- EXTERNAL_ATTENUATION = 4194307
Specifies the attenuation of a switch or cable connected to the RF IN connector of the signal analyzer. This value is expressed in dB. For more information about attenuation, refer to the RF Attenuation and Signal Levels topic for your device in the NI RF Vector Signal Analyzers Help.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.
- HB_FILTER_WIDTH = 4194322
Specifies the filter width when you set the
BURST_TYPEattribute to HB.Use “slot<n>” as the selector string to configure or read this attribute.
The default value is Narrow.
Name (Value)
Description
Narrow (0)
The measurement uses a narrow filter.
Wide (1)
The measurement uses a wide filter.
- IQ_POWER_EDGE_TRIGGER_LEVEL = 4194312
Specifies the power level at which the device triggers. This value is expressed in dB when you set the
IQ_POWER_EDGE_TRIGGER_LEVEL_TYPEattribute to Relative and in dBm when you set the IQ Power Edge Level Type attribute to Absolute. The device asserts the trigger when the signal exceeds the level specified by the value of this attribute, taking into consideration the specified slope. This attribute is used only when you set theTRIGGER_TYPEattribute to IQ Power Edge.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default of this attribute is hardware dependent.
- IQ_POWER_EDGE_TRIGGER_LEVEL_TYPE = 4198399
Specifies the reference for the
IQ_POWER_EDGE_TRIGGER_LEVELattribute. The IQ Power Edge Level Type attribute is used only when you set theTRIGGER_TYPEattribute to IQ Power Edge.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Relative.
Name (Value)
Description
Relative (0)
The value of the IQ Power Edge Level attribute is relative to the value of the Reference Level attribute.
Absolute (1)
The IQ Power Edge Level attribute specifies the absolute power.
- IQ_POWER_EDGE_TRIGGER_SLOPE = 4194313
Specifies whether the device asserts the trigger when the signal power is rising or when it is falling. The device asserts the trigger when the signal power exceeds the level that you specify in the
IQ_POWER_EDGE_TRIGGER_LEVELattribute with the slope you specify. This attribute is used only when you set theTRIGGER_TYPEattribute to IQ Power Edge.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Rising Slope.
Name (Value)
Description
Rising Slope (0)
The trigger asserts when the signal power is rising.
Falling Slope (1)
The trigger asserts when the signal power is falling.
- IQ_POWER_EDGE_TRIGGER_SOURCE = 4194311
Specifies the channel from which the device monitors the trigger. This attribute is used only when you set the
TRIGGER_TYPEattribute to IQ Power Edge.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default of this attribute is hardware dependent.
- LIMITED_CONFIGURATION_CHANGE = 4247555
Specifies the set of attributes that are considered by RFmx in the locked signal configuration state.
If your test system performs the same measurement at different selected ports, multiple frequencies and/or power levels repeatedly, enabling this attribute can help achieve faster measurements. When you set this attribute to a value other than Disabled, the RFmx driver will use an optimized code path and skip some checks. Because RFmx skips some checks when you use this attribute, you need to be aware of the limitations of this feature, which are listed in the Limitations of the Limited Configuration Change Property topic.
You can also use this attribute to lock a specific instrument configuration for a signal so that every time that you initiate the signal, RFmx applies the RFmxInstr attributes from a locked configuration.
NI recommends you use this attribute in conjunction with named signal configurations. Create named signal configurations for each measurement configuration in your test program and set this attribute to a value other than Disabled for one or more of the named signal configurations. This allows RFmx to precompute the acquisition settings for your measurement configurations and re-use the precomputed settings each time you initiate the measurement. You do not need to use this attribute if you create named signals for all the measurement configurations in your test program during test sequence initialization and do not change any RFInstr or personality attributes while testing each device under test. RFmx automatically optimizes that use case.
Specify the named signal configuration you are setting this attribute in the selector string input. You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Disabled.
Name (Value)
Description
Disabled (0)
This is the normal mode of RFmx operation. All configuration changes in RFmxInstr attributes or in personality attributes will be applied during RFmx Commit.
No Change (1)
Signal configuration is locked after the first Commit of the named signal configuration. Any configuration change thereafter either in RFmxInstr attributes or personality attributes will not be considered by subsequent RFmx Commits or Initiates of this signal. Use No Change if you have created named signal configurations for all measurement configurations but are setting some RFmxInstr attributes. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode.
Frequency (2)
Signal configuration, other than center frequency and external attenuation, is locked after first Commit of the named signal configuration. Thereafter, only the Center Frequency and External Attenuation attribute value changes will be considered by subsequent driver Commits or Initiates of this signal. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode.
Reference Level (3)
Signal configuration, other than the reference level, is locked after first Commit of the named signal configuration. Thereafter only the Reference Level attribute value change will be considered by subsequent driver Commits or Initiates of this signal. If you have configured this signal to use an IQ Power Edge Trigger, NI recommends that you set the IQ Power Edge Level Type to Relative so that the trigger level is automatically adjusted as you adjust the reference level. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode.
Freq and Ref Level (4)
Signal configuration, other than center frequency, reference level, and external attenuation, is locked after first Commit of the named signal configuration. Thereafter only Center Frequency, Reference Level, and External Attenuation attribute value changes will be considered by subsequent driver Commits or Initiates of this signal. If you have configured this signal to use an IQ Power Edge Trigger, NI recommends you set the IQ Power Edge Level Type to Relative so that the trigger level is automatically adjusted as you adjust the reference level. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode.
Selected Ports, Freq and Ref Level (5)
Signal configuration, other than selected ports, center frequency, reference level, external attenuation, and RFInstr configuration, is locked after first Commit or Initiate of the named signal configuration. Thereafter only Selected Ports, Center Frequency, Reference Level, and External Attenuation attribute value changes will be considered by subsequent driver Commits or Initiates of this signal. If you have configured this signal to use an IQ Power Edge Trigger, NI recommends you set the IQ Power Edge Level Type to Relative so that the trigger level is automatically adjusted as you adjust the reference level. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode.
- LINK_DIRECTION = 4194317
Specifies the source of the signal to be measured.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Uplink.
Name (Value)
Description
Downlink (0)
The source is a base transceiver station.
Uplink (1)
The source is a mobile station.
- MODACC_ALL_TRACES_ENABLED = 4198406
Specifies whether to enable the traces to be stored and retrieved after performing the modulation accuracy (ModAcc) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- MODACC_AVERAGING_COUNT = 4198404
Specifies the number of acquisitions used for averaging when you set the
MODACC_AVERAGING_ENABLEDattribute to True.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10.
- MODACC_AVERAGING_ENABLED = 4198402
Specifies whether to enable averaging for the modulation accuracy (ModAcc) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
The measurement is performed on a single acquisition.
True (1)
The measurement is averaged over multiple acquisitions.
- MODACC_DROOP_COMPENSATION_ENABLED = 4198405
Specifies whether to enable droop compensation for the modulation accuracy (ModAcc) measurement. Droop compensation allows the ModAcc measurement to minimize the contribution of amplifier power variations to the EVM results.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
Disables power droop compensation in the EVM measurement.
True (1)
Enables power droop compensation in the EVM measurement.
- MODACC_MEASUREMENT_ENABLED = 4198400
Specifies whether to enable modulation accuracy (ModAcc) measurements on the acquired signal.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- MODACC_MEASUREMENT_INTERVAL = 4198442
Specifies whether the measurement is performed on a single specified timeslot or across multiple timeslots.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default Value is ‘Number of Timeslots’.
Name (Value)
Description
Number of Timeslots (0)
The measurement includes all timeslots defined by the Num Timeslots attribute.
Timeslot at Offset (1)
The measurement is performed only on the timeslot specified by the ModAcc Meas Offset attribute. This attribute is applicable only when you set the Trigger Type attribute to IQ Power Edge or **Digital Edge **.
- MODACC_MEASUREMENT_OFFSET = 4198443
Specifies the offset, relative to the trigger of the timeslot to be measured.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0. Valid values are [0, (
NUMBER_OF_TIMESLOTS-1)].
- MODACC_NUMBER_OF_ANALYSIS_THREADS = 4198407
Specifies the maximum number of threads used for parallelism for the ModAcc measurement.
The number of threads can range from 1 to the number of physical cores. The number of threads you set may not be used in calculations. The actual number of threads used depends on the problem size, system resources, data availability, and other considerations.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 1.
- MODACC_RESULTS_DETECTED_TSC = 4198433
Returns the detected training sequence code (TSC) if you set the
BURST_SYNCHRONIZATION_TYPEattribute to TSC.Use “slot<n>” as the selector string to read this result.
When the
MODACC_MEASUREMENT_INTERVALattribute is set to Timeslot at Offset, Detected TSC array has one element and the Detected TSC is returned at index 0.Name (Value)
Description
Unknown (-1)
The synchronization is not found, and measurements correspond to best estimate of synchronization.
TSC0 (0)
The detected TSC is TSC0.
TSC1 (1)
The detected TSC is TSC1.
TSC2 (2)
The detected TSC is TSC2.
TSC3 (3)
The detected TSC is TSC3.
TSC4 (4)
The detected TSC is TSC4.
TSC5 (5)
The detected TSC is TSC5.
TSC6 (6)
The detected TSC is TSC6.
TSC7 (7)
The detected TSC is TSC7.
- MODACC_RESULTS_EVM_95TH_PERCENTILE_EVM = 4198412
Returns the EVM value at which no more than 5% of the symbols have an EVM exceeding this value. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_95TH_PERCENTILE_MAGNITUDE_ERROR = 4198437
Returns the measured magnitude error value multiplied by 100, at which, no more than 5 percent of the symbols have magnitude error exceeding this value. This result is returned while performing ModAcc measurement on EDGE/EGPRS. This value is measured as a percentage.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_95TH_PERCENTILE_PHASE_ERROR = 4198440
Returns the measured phase error value multiplied by 100, at which, no more than 5 percent of the symbols have phase error exceeding this value. This result is returned while performing ModAcc measurement on EDGE/EGPRS. This value is measured in degrees.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_AMPLITUDE_DROOP = 4198421
Returns the maximum of the amplitude droop values over all the measured time slots. This value is expressed in dB/symbol. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_FREQUENCY_ERROR = 4198419
Returns the maximum of the frequency error values over all the measured time slots. This value is expressed in Hz. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_MAGNITUDE_ERROR = 4198415
Returns the maximum of RMS values of magnitude error over all the measured time slots. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_PEAK_EVM = 4198411
Returns the maximum of the peak EVM values over all the measured time slots. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_PEAK_MAGNITUDE_ERROR = 4198436
Returns the maximum of peak magnitude error values over all the measured time slots. This result is returned while performing ModAcc measurement on EDGE/EGPRS. This value is measured as a percentage.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_PEAK_PHASE_ERROR = 4198439
Returns the maximum of peak phase error values over all the measured time slots. This result is returned while performing ModAcc measurement on EDGE/EGPRS. This value is measured in degrees.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_PHASE_ERROR = 4198417
Returns the maximum of the RMS values of phase error over all the measured time slots. This value is expressed in degrees. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MAXIMUM_RMS_EVM = 4198409
Returns the maximum of RMS values of EVM over all the measured time slots. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_AMPLITUDE_DROOP = 4198420
Returns the mean of the amplitude droop values over all the measured time slots. This value is expressed in dB/symbol. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_FREQUENCY_ERROR = 4198418
Returns the mean of the frequency error values over all the measured time slots. This value is expressed in Hz. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_MAGNITUDE_ERROR = 4198414
Returns the mean of RMS values of magnitude error over all the measured time slots. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_PEAK_EVM = 4198410
Returns the mean of the peak EVM values over all the measured time slots. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_PEAK_MAGNITUDE_ERROR = 4198435
Returns the mean of peak magnitude error values over all the measured time slots. This result is returned while performing ModAcc measurement on EDGE/EGPRS. This value is measured as a percentage.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_PEAK_PHASE_ERROR = 4198438
Returns the mean of peak phase error values over all the measured time slots. This result is returned while performing ModAcc measurement on EDGE/EGPRS. This value is measured in degrees.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results. &gt;
- MODACC_RESULTS_EVM_MEAN_PHASE_ERROR = 4198416
Returns the mean of RMS values of phase error over all the measured time slots. This value is expressed in degrees. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_MEAN_RMS_EVM = 4198408
Returns the mean of RMS values of EVM over all the measured time slots. This value is expressed as a percentage. The attribute returns this result for EDGE/EGPRS measurements.
You do not need to use a Selector String to configure or read this attribute for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_EVM_PEAK_EVM_SYMBOL = 4198413
Returns the symbol number in the useful portion of the burst corresponding to
MODACC_RESULTS_EVM_MAXIMUM_PEAK_EVMresult. The attribute returns this result for ModAcc EDGE/EGPRS measurements.You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_MAXIMUM_IQ_GAIN_IMBALANCE = 4198430
Returns the maximum of I/Q gain imbalance values over all the measured time slots. This value is expressed in dB. The presence of quadrature skew in the signal affects the measurement of I/Q gain imbalance.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_MAXIMUM_IQ_ORIGIN_OFFSET = 4198432
Returns the maximum of the I/Q origin offset values over all the measured time slots. This value is expressed in dB. Presence of I/Q gain imbalance and quadrature skew in the signal affects the I/Q origin offset measurement. The attribute returns this result for GSM/EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_MEAN_IQ_GAIN_IMBALANCE = 4198429
Returns the mean of I/Q gain imbalance values over all the measured time slots. This value is expressed in dB. The presence of quadrature skew in the signal affects the measurement of I/Q gain imbalance.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_MEAN_IQ_ORIGIN_OFFSET = 4198431
Returns the mean of the I/Q origin offset values over all the measured time slots. This value is expressed in dB. Presence of I/Q gain imbalance and quadrature skew in the signal affects the I/Q origin offset measurement. The attribute returns this result for GSM/EDGE/EGPRS measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_95TH_PERCENTILE_PHASE_ERROR = 4198441
Returns the measured phase error value multiplied by 100, at which, no more than 5 percent of the symbols have phase error exceeding this value. This result is returned while performing ModAcc measurement on GSM. This value is measured in degrees.
- MODACC_RESULTS_PFER_MAXIMUM_FREQUENCY_ERROR = 4198428
Returns the maximum of the frequency error values over all the measured time slots. This value is expressed in Hz. The attribute returns this result for GSM ModAcc measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_MAXIMUM_PEAK_PHASE_ERROR = 4198425
Returns the maximum of peak phase error values over all the measured time slots. This value is expressed in degrees. The attribute returns this result for GSM ModAcc measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_MAXIMUM_RMS_PHASE_ERROR = 4198423
Returns the maximum of the RMS values of phase error over all the measured time slots. This value is expressed in degrees. The attribute returns this result for GSM ModAcc measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_MEAN_FREQUENCY_ERROR = 4198427
Returns the mean of the frequency error values over all the measured time slots. This value is expressed in Hz. The attribute returns this result for GSM ModAcc measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_MEAN_PEAK_PHASE_ERROR = 4198424
Returns the mean of peak phase error values over all the measured time slots. This value is expressed in degrees. The attribute returns this result for GSM ModAcc measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_MEAN_RMS_PHASE_ERROR = 4198422
Returns the mean of RMS values of phase error over all the measured time slots. This value is expressed in degrees. The attribute returns this result for GSM ModAcc measurements.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODACC_RESULTS_PFER_PEAK_PHASE_ERROR_SYMBOL = 4198426
Returns the symbol number in the useful portion of the burst corresponding to the phase error value in the
MODACC_RESULTS_PFER_MAXIMUM_PEAK_PHASE_ERRORresult. The attribute returns this result for GSM ModAcc measurements.You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
- MODULATION_TYPE = 4194320
Specifies the modulation scheme used for the signal.
Use “slot<n>” as the selector string to configure or read this attribute.
The default value is 8PSK.
Name (Value)
Description
GMSK (0)
The modulation type is Gaussian minimum shift keying. This value is valid only when you set the Burst Type attribute to NB or AB.
8PSK (1)
The modulation type is 8-PSK. This value is valid only when you set the Burst Type attribute to NB.
QPSK (2)
The modulation type is QPSK. This value is valid only when you set the Burst Type attribute to HB.
16QAM (3)
The modulation type is 16-QAM.
32QAM (4)
The modulation type is 32-QAM.
- NUMBER_OF_TIMESLOTS = 4194319
Specifies the number of time slots to be measured.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 1.
- ORFS_ALL_TRACES_ENABLED = 4202516
Specifies whether to enable the traces to be stored and retrieved after performing the output radio frequency spectrum (ORFS) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- ORFS_AVERAGING_COUNT = 4202500
Specifies the number of acquisitions used for averaging when you set the
ORFS_AVERAGING_ENABLEDattribute to True.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10.
- ORFS_AVERAGING_ENABLED = 4202498
Specifies whether to enable averaging for the output radio frequency spectrum (ORFS) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
The measurement is performed on a single acquisition.
True (1)
The measurement is averaged over multiple acquisitions.
- ORFS_AVERAGING_TYPE = 4202499
Specifies the averaging type for averaging multiple spectrum acquisitions. The averaged spectrum is used for the output radio frequency spectrum (ORFS) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Log.
Name (Value)
Description
RMS (0)
The measurement averages the power spectrum linearly. RMS averaging reduces signal fluctuations but not the noise floor.
Log (1)
The measurement averages the power spectrum in a logarithmic scale.
- ORFS_EVALUATION_SYMBOLS_INCLUDE_TSC = 4202514
Specifies whether to include the training sequence code (TSC) portion of burst for measuring ORFS due to modulation.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
The TSC portion of the burst will be excluded while performing the ORFS measurement.
True (1)
The TSC portion of the burst will be included while performing the ORFS measurement.
- ORFS_EVALUATION_SYMBOLS_SCOPE = 4202532
Specifies the modulation power measurements that will use the part of burst configured using the
configure_evaluation_symbols().You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Offset.
Name (Value)
Description
Offset (0)
Modulation Offset Power measurements will only use the part of burst configured using the RFmxGSM ORFS Configure Evaluation Symbols method. Modulation Carrier Power measurement will use all of the useful part of the burst.
Offset and Carrier (1)
Modulation Offset and Carrier Power measurements will both use the part of burst configured using the RFmxGSM ORFS Configure Evaluation Symbols method.
- ORFS_EVALUATION_SYMBOLS_START = 4202513
Specifies the start position within the useful part of the burst, in percentage, for measuring ORFS due to modulation.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 50.
- ORFS_EVALUATION_SYMBOLS_STOP = 4202515
Specifies the stop position within the useful part of the burst, in percentage, for measuring ORFS due to modulation.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 90.
- ORFS_MEASUREMENT_ENABLED = 4202496
Specifies whether to enable the output radio frequency spectrum (ORFS) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- ORFS_MEASUREMENT_INTERVAL = 4202533
Specifies whether the measurement is performed on a single specified timeslot or across multiple timeslots.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default Value is ‘Number of Timeslots’.
Name (Value)
Description
Number of Timeslots (0)
The measurement includes all timeslots defined by the Num Timeslots attribute.
Timeslot at Offset (1)
The measurement is performed only on the timeslot specified by the ORFS Meas Offset attribute. This attribute is applicable only when you set the Trigger Type attribute to IQ Power Edge or **Digital Edge **.
- ORFS_MEASUREMENT_OFFSET = 4202534
Specifies the offset, relative to the trigger, of the timeslot to be measured.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0. Valid values are [0, (
NUMBER_OF_TIMESLOTS-1)].
- ORFS_MEASUREMENT_TYPE = 4202501
Specifies the type of spectral distortion to be measured.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Modulation and Switching.
Name (Value)
Description
Modulation and Switching (0)
Measures the spectrum on the modulated part and the switching part of the waveform.
Modulation (1)
Measures the spectrum on the modulated part of the waveform.
Switching (2)
Measures the spectrum on the switching part of the waveform.
- ORFS_MODULATION_CARRIER_RBW = 4202507
Specifies the RBW used for measuring modulation carrier power in Hz.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 30kHz.
- ORFS_MODULATION_NUMBER_OF_OFFSETS = 4202525
Specifies the number of positive frequency offsets relative to the frequency of the carrier for the measurement of the spectrum due to modulation.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.
- ORFS_MODULATION_OFFSET_FREQUENCY = 4202526
Specifies the value of positive frequency offset for which to measure the spectrum due to modulation measurement. This value is expressed in Hz.
Use “offset<n>” as the selector string to configure or read this attribute.
The default value is 0.
- ORFS_MODULATION_OFFSET_RBW = 4202527
Specifies the resolution bandwidth used for ORFS due to modulation measurement. This value is expressed in Hz.
Use “offset<n>” as the selector string to configure or read this attribute.
The default value is 30000.
- ORFS_NOISE_COMPENSATION_ENABLED = 4202502
Specifies whether to enable compensation of the channel powers for the inherent noise floor of the signal analyzer. Noise compensation is applicable only on modulation offsets and not on switching offsets.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Supported Devices: PXIe-5663/5665/5668R, PXIe-5830/5831/5832
Name (Value)
Description
False (0)
Disables compensation of the channel powers for the signal analyzer noise floor.
True (1)
Enables compensation of the channel powers for the noise floor of the signal analyzer. The noise floor of the signal analyzer is measured for the RF path used by the output radio frequency spectrum (ORFS) measurement and cached for future use. If signal analyzer or measurement parameters change, noise floors are measured again.
- ORFS_NUMBER_OF_ANALYSIS_THREADS = 4202517
Specifies the maximum number of threads used for parallelism for the output radio frequency spectrum (ORFS) measurement.
The number of threads can range from 1 to the number of physical cores. The number of threads you set may not be used in calculations. The actual number of threads used depends on the problem size, system resources, data availability, and other considerations.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 1.
- ORFS_OFFSET_FREQUENCY_MODE = 4202503
Specifies the list of frequency offsets for which you can perform the output radio frequency spectrum (ORFS) measurements.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Standard.
Name (Value)
Description
Standard (0)
Uses the list of lower and upper offset frequencies of 100 kHz, 200 kHz, 250 kHz, 400 kHz, 600 kHz, 800 kHz, 1000 kHz, 1200 kHz, 1400 kHz, 1600 kHz, and 1800 kHz for the spectrum due to the modulation measurement, and the lower and upper offset frequencies of 400 kHz, 600 kHz, 1200 kHz, and 1800 kHz for the spectrum produced by the switching measurement.
Short (1)
Uses the list of lower and upper offset frequencies of 200 kHz, 250 kHz, 400 kHz, 600 kHz, and 1200 kHz for the spectrum due to the modulation measurement, and the lower and upper offset frequencies of 400 kHz, 600 kHz, 1200 kHz, and 1800 kHz for the spectrum produced by the switching measurement.
Custom (2)
Uses the list of frequencies that is configured using the RFmxGSM ORFS Configure Modulation Custom Offset Frequency (Array) and RFmxGSM ORFS Configure Switching Custom Offset Frequency (Array) methods for measurement of spectrum produced by modulation and switching measurements.
- ORFS_SWITCHING_CARRIER_RBW = 4202510
Specifies the RBW used for measuring switching carrier power in Hz.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 300kHz.
- ORFS_SWITCHING_NUMBER_OF_OFFSETS = 4202528
Specifies the number of positive frequency offsets relative to the frequency of the carrier for the measurement of the spectrum due to switching.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.
- ORFS_SWITCHING_OFFSET_FREQUENCY = 4202529
Specifies the value of positive frequency offset for which to measure the spectrum due to switching measurement. This value is expressed in Hz.
Use “offset<n>” as the selector string to configure or read this attribute.
The default value is 0.
- ORFS_SWITCHING_OFFSET_RBW = 4202530
Specifies the resolution bandwidth used for ORFS due to switching measurement. This value is expressed in Hz.
Use “offset<n>” as the selector string to configure or read this attribute.
The default value is 30kHz.
- POWER_CONTROL_LEVEL = 4194325
Specifies the power control level corresponding to the transmitted power, as defined in section 4.1 of the 3GPP TS 45.005 v8.0.0 specifications.
Use “slot<n>” as the selector string to configure or read this attribute.
The default value is 0. Valid values are 0 to 31, inclusive.
- PVT_ALL_TRACES_ENABLED = 4206601
Specifies whether to enable the traces to be stored and retrieved after performing the power versus time (PVT) measurement.
The default value is FALSE.
- PVT_AVERAGING_COUNT = 4206597
Specifies the number of acquisitions used for averaging when you set the
PVT_AVERAGING_ENABLEDattribute to True.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10.
- PVT_AVERAGING_ENABLED = 4206594
Specifies whether to enable averaging for the power versus time (PVT) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is False.
Name (Value)
Description
False (0)
The measurement is performed on a single acquisition.
True (1)
The measurement is averaged over multiple acquisitions.
- PVT_AVERAGING_TYPE = 4206596
Specifies the averaging type for multiple acquisitions.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is RMS.
Name (Value)
Description
RMS (0)
The power at each sample interval is linearly averaged from one acquisition to the next acquisition.
Log (1)
The power at each sample interval is averaged on a logarithmic scale from one acquisition to the next acquisition.
Max (3)
The peak power at each sample interval is retained from one acquisition to the next acquisition.
Min (4)
The lowest power at each sample interval is retained from one acquisition to the next acquisition.
- PVT_MAXIMUM_POWER_TRACE_ENABLED = 4206611
Specifies whether the maximum power trace is enabled. When
PVT_AVERAGING_ENABLEDattribute is set to True, the peak power at each sample interval is retained from one acquisition to the next acquisition. You can use the RFmxGSM PVT Fetch Maximum Power Trace method to fetch the maximum power trace.This attribute is applicable only when you set the
PVT_AVERAGING_TYPEattribute to RMS or Log.The default value is FALSE.
Name (Value)
Description
False (0)
Disables computation of the PVT Maximum Power Trace.
True (1)
Enables computation of the PVT Maximum Power Trace.
- PVT_MEASUREMENT_ENABLED = 4206592
Specifies whether to enable the power versus time (PVT) measurement.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is FALSE.
- PVT_MINIMUM_POWER_TRACE_ENABLED = 4206612
Specifies whether the minimum power trace is enabled. When
PVT_AVERAGING_ENABLEDattribute is set to True, the lowest power at each sample interval is retained from one acquisition to the next acquisition. You can use the RFmxGSM PVT Fetch Minimum Power Trace method to fetch the minimum power trace.This attribute is applicable only when you set the
PVT_AVERAGING_TYPEattribute to RMS or Log.The default value is FALSE.
Name (Value)
Description
False (0)
Disables computation of the PVT Minimum Power Trace.
True (1)
Enables computation of the PVT Minimum Power Trace.
- PVT_NUMBER_OF_ANALYSIS_THREADS = 4206602
Specifies the maximum number of threads used for parallelism for the power versus time (PVT) measurement.
The number of threads can range from 1 to the number of physical cores. The number of threads you set may not be used in calculations. The actual number of threads used depends on the problem size, system resources, data availability, and other considerations.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 1.
- PVT_RBW_FILTER_BANDWIDTH = 4206599
Specifies the RBW Filter Bandwidth in Hz
The default value is 500 kHz
- PVT_RESULTS_MEASUREMENT_STATUS = 4206603
Indicates the overall measurement status based on standard-defined limits.
You do not need to use a Selector String to read this result for the default signal and result instances. Refer to the Selector String topic for information about the string syntax for named signals and named results.
Name (Value)
Description
Fail (0)
The measurement failed because the average power of at least one slot is outside the standard-defined limits.
Pass (1)
The measurement passed because the average power of all slots is within the standard-defined limits.
- PVT_RESULTS_SLOT_AVERAGE_POWER = 4206604
Returns the mean power of the signal, averaged over the corresponding slots of each acquisition. This value is expressed in dBm.
Use “slot<n>” as the selector string to read this attribute.
- PVT_RESULTS_SLOT_BURST_THRESHOLD = 4206608
Returns the threshold that the power versus time (PVT) measurement uses to determine the burst validity. This value is expressed in dBm.
Use “slot<n>” as the selector string to read this attribute.
- PVT_RESULTS_SLOT_BURST_WIDTH = 4206605
Returns the burst width for the slot where the -3 dB transition points occur. This value is expressed in seconds.
Use “slot<n>” as the selector string to read this attribute.
- PVT_RESULTS_SLOT_MAXIMUM_POWER = 4206606
Returns the maximum power of the signal, averaged over the corresponding slots of each acquisition. This value is expressed in dBm.
Use “slot<n>” as the selector string to read this attribute.
- PVT_RESULTS_SLOT_MEASUREMENT_STATUS = 4206609
Indicates the power versus time (PVT) measurement status for a particular slot.
Use “slot<n>” as the selector string to read this attribute.
Name (Value)
Description
Fail (0)
The average power for at least one slot is outside the standard-defined limits.
Pass (1)
The average power for each slot is within the standard-defined limits.
- PVT_RESULTS_SLOT_MINIMUM_POWER = 4206607
Returns the minimum power of the signal, averaged over the corresponding slots of each acquisition. This value is expressed in dBm.
Use “slot<n>” as the selector string to read this attribute.
- REFERENCE_LEVEL = 4194306
Specifies the reference level that represents the maximum expected power of the RF input signal. This value is expressed in dBm for RF devices and Vpk-pkfor baseband devices.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default of this attribute is hardware dependent.
- REFERENCE_LEVEL_HEADROOM = 4198396
Specifies the margin RFmx adds to the
REFERENCE_LEVELattribute. The margin avoids clipping and overflow warnings if the input signal exceeds the configured reference level.RFmx configures the input gain to avoid clipping and associated overflow warnings provided the instantaneous power of the input signal remains within the Reference Level plus the Reference Level Headroom. If you know the input power of the signal precisely or previously included the margin in the Reference Level, you could improve the signal-to-noise ratio by reducing the Reference Level Headroom.
You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
**Supported devices: **PXIe-5668R, PXIe-5830/5831/5832/5840/5841/5842/5860.
Default values
Name (value)
Description
PXIe-5668
6 dB
PXIe-5830/5831/5832/5841/5842/5860
1 dB
PXIe-5840
0 dB
- RESULT_FETCH_TIMEOUT = 4243456
Specifies the time to wait before results are available in the RFmxGSM Attribute. This value is expressed in seconds.
Set this value to a time longer than expected for fetching the measurement. A value of -1 specifies that the RFmxGSM Attribute waits until the measurement is complete.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 10.
- SELECTED_PORTS = 4198397
Specifies the instrument port to be configured to acquire a signal. Use
nirfmxinstr.session.Session.get_available_ports()method to get the valid port names.You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
Valid values
Name (value)
Description
PXIe-5830
if0, if1
PXIe-5831/5832
if0, if1, rf<0-1>/port<x>, where 0-1 indicates one (0) or two (1) mmRH-5582 connections and x is the port number on the mmRH-5582 front panel
Other devices
“” (empty string)
Default values
Name (value)
Description
PXIe-5830/5831/5832
if1
Other devices
“” (empty string)
- SIGNAL_STRUCTURE = 4194327
Specifies whether the signal is bursted or continuous. For bursted signal and continuous signals, set the
TRIGGER_TYPEto IQ Power Edge and None, respectively.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Bursted.
Name (Value)
Description
Bursted (0)
The signal is bursted.
Continuous (1)
The signal is continuous.
- TIMING_ADVANCE = 4194328
Specifies the timing advance value as specified in the 3GPP TS 45.010 specification for GSM access burst.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.
- TRIGGER_DELAY = 4194314
Specifies the trigger delay time. This value is expressed in seconds.
If the delay is negative, the measurement acquires pretrigger samples. If the delay is positive, the measurement acquires post-trigger samples.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is 0.
- TRIGGER_MINIMUM_QUIET_TIME_DURATION = 4194316
Specifies the time duration for which the signal must be quiet before the signal analyzer arms the I/Q power edge trigger. This value is expressed in seconds.
If you set the
IQ_POWER_EDGE_TRIGGER_SLOPEattribute to Rising Slope, the signal is quiet below the trigger level. If you set the IQ Power Edge Slope attribute to Falling Slope, the signal is quiet above the trigger level.You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default of this attribute is hardware dependent.
- TRIGGER_MINIMUM_QUIET_TIME_MODE = 4194315
Specifies whether the measurement computes the minimum quiet time used for triggering.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Auto.
Name (Value)
Description
Manual (0)
The minimum quiet time for triggering is the value of the Trigger Min Quiet Time attribute.
Auto (1)
The measurement computes the minimum quiet time used for triggering.
- TRIGGER_TYPE = 4194308
Specifies the trigger type.
You do not need to use a Selector String to configure or read this attribute for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is IQ Power Edge.
Name (Value)
Description
None (0)
No Reference Trigger is configured.
Digital Edge (1)
The Reference Trigger is not asserted until a digital edge is detected. The source of the digital edge is specified using the Digital Edge Source attribute.
IQ Power Edge (2)
The Reference Trigger is asserted when the signal changes past the level specified by the slope (rising or falling), which is configured using the IQ Power Edge Slope attribute.
Software (3)
The Reference Trigger is not asserted until a software trigger occurs.
- TSC = 4194324
Specifies the training sequence code (TSC) to use. This attribute is applicable only when you set the
BURST_SYNCHRONIZATION_TYPEattribute to TSC and theAUTO_TSC_DETECTION_ENABLEDattribute to False. For access burst TSC0, TSC1, and TSC2 are applicable.Use “slot<n>” as the selector string to configure or read this attribute.
The default value is TSC0.
Name (Value)
Description
TSC0 (0)
The measurement uses training sequence code 0.
TSC1 (1)
The measurement uses training sequence code 1.
TSC2 (2)
The measurement uses training sequence code 2.
TSC3 (3)
The measurement uses training sequence code 3.
TSC4 (4)
The measurement uses training sequence code 4.
TSC5 (5)
The measurement uses training sequence code 5.
TSC6 (6)
The measurement uses training sequence code 6.
TSC7 (7)
The measurement uses training sequence code 7.